- Accuracy, RH: RHP-2XXX ±2% 10-90% RH @ 25◦C;
- RHP-3XXX ±3% 20-80% RH @ 25◦C;
- RHP-5XXX ±5% 20-80% RH @ 25◦C.
- Accuracy, Thermistor Temp Sensor: ±0.2◦C @ 25◦C (±0.36◦F@77◦F)
- Accuracy, RTD Temp Sensor: DIN Class B; ±0.3◦C@0◦C (±0.54◦F@32◦F)
- Accuracy, Solid State Band Gap: ±0.9◦F@ 77◦F (±0.3◦C@25◦C)
- Hysteresis: ±1%
- Repeatability: ±0.1% Typical
- Temperature Limit: -40 to 140â—¦F(-40 to 60â—¦C)
- Storage Temperature: -40 to 176â—¦F(-40 to 80â—¦C)
- Compensated Temperature Range: -4 to 140â—¦F(-20 to 60â—¦C)
- 4 to 20mA Loop Powered Models: Power requirements: 10 to 35 VDC; Output signal:4 to 20mA
- 0-5/10V Output Models: Power requirements: 15 to 35 VDC or 15 to 29VAC; Output signal: 0to 10V@ 5mA max.
- Solid State Band Gap Temperature Sensor Output Ranges: Switch selectable, -20 to 140â—¦F (-28.9 to 60â—¦C);0 to 100â—¦F(-17.8 to 37.8â—¦C); 40 to 90â—¦F (4.4 to 32.3â—¦C);-4 to 140â—¦F (-20 to 60â—¦C)
- Electrical Connection: Removable screw terminal block
- Conduit Connection: Duct mount: ½â€NPS; OSA: ½†(22.3mm)
- Drift:
Place of Origin | India |
Minimum Order Quantity | Not Given |
Supply Ability | Not Given |
Packaging Details | Not Given |
Delivery Details | Not Given |
Not Given